Products

AIJGO-34 Personal grounding tester

The AIJGO-34 is a tester for checking personal grounding before entering an ESD protected area (EPA). It may be a good choice if you want to identify persons entering the site. Tester with LIGHT BAR system – such AIJGO products have a steel touch button, the measured results are indicated by LED lights. COMFORT system product – such AIJGO devices have touch button and RFID reader, an RFID card is required for use, which includes the type of test to be used.

The tester is supplied with manufacturer’s calibration  certificate, dual foot electrode, mounting plate and universal holder.

Download Datasheet

Download User Manual

Download PieManager User Manual


AIJGO-34 PERSONAL GROUNDING TESTER

WITH DUAL FOOT ELECTRODE, RFID READER, LCD DISPLAY AND OPERATING SOFTWARE

ABOUT THE PRODUCT

Aijgo 34
  
  • Tester for checking personal grounding before entering an ESD protected area (EPA)
  • Testing footwear system and wrist strap/groundable ESD smock system at the same time
  • Can be used to test a groundable ESD smock system
  • Test result indication on LCD display, with LED light and sound
  • For indoor use only
Light BarCONFORT system

 

 

The AIJGO-34 is a tester for checking personal grounding before entering an ESD protected area (EPA). It may be a good choice if you want to identify persons entering the site. Tester with LIGHT BAR system – such AIJGO products have a steel touch button, the measured results are indicated by LED lights. COMFORT system product – such AIJGO devices have touch button and RFID reader, an RFID card is required for use, which includes the type of test to be used.

This type can be:

  1. Footwear system test
  2. Wrist strap/groundable ESD smock system test
  3. Footwear system and wrist strap/groundable ESD smock system combined test
  4. VIP test. The test can start after the card has been scanned.

Thanks to the dual foot electrode, the tester simultaneously measures the resistance of the right and left foot relative to the hand. After the test, the measured values are displayed on the built-in LCD display. The measured results are stored on an internal SD card. The tester can be connected to a gate using the relay output (e.g. to a turnstile entrance gate, swing gate, etc.). It can be installed on an entrance gate, table, wall or stand. It is also possible to configure it so that a test can be carried out without touching the touch button, but an RFID card is required in this case too. The product is supplied with an operating software and can operate individually with this. If required, a management software called PieManager can be purchased to configure it over a LAN network and to retrieve the measured results. The system assigns a test method to users and manages their identification data. It saves the entry data as .csv files (compatible with Excel and text editor) and automatically logs daily events.

It is recommended and sufficient to purchase only one management software if several testers are purchased and networked. Purchasing and networking multiple management software with multiple testers can lead to communication and system malfunctions. If you have several testers and you do not network them, but use them as separate devices, please buy individual software for each device to configure and collect data over the LAN network.

The tester is supplied with manufacturer’s calibration certificate, dual foot electrode, mounting plate and universal holder.

Aijgo 34

aijgo 34

aijgo 34

CE declaration

We declare that the AIJGO-34 product complies with the requirements of IEC 61340-5-1, ANSI/ESD S20.20 and Directive 2001/95/EC (General product safety).

Warning

The device can deliver an effective output power of up to 20 µA or less at a maximum voltage of 100 V for 2-4 seconds at 0,5 second cycles. Therefore, persons with an implanted electronic device (e.g. insulin pump, pacemaker, etc.), known or acute hear rhythm disorders and other disorders of the heart’s impulse generation and conduction, seizure disorders (e.g. epilepsy) should consult their doctor before using this device.

TECHNICAL INFORMATIONS

Resistance values indicated as acceptable by the tester

Default setting when measuring footwear system 100 KΩ – 100 MΩ
You can also set an upper value of 35 MΩ, see the PieManager management software description.
Default setting when measuring wrist strap/
groundable ESD smock system
750 KΩ – 35 MΩ

 

Operating system Linux Raspbian
The tester can be configured via LAN network using RJ45 cable
Test voltage 100V DC
Wrist strap/groundable ESD smock system connection With a combined 10 mm snap, 4 mm banana plug socket
Power supply 12 V DC
Relay outputs 0,5 A/60 V
Pass relay (Normally Open) with green wire
Fail relay (Normally Open) with red wire
Sizes 300 x 195 x 55 mm (tester)
500 x 630 x 970 mm (tester+stand with dual foot electrode, assembled), the touch button and the RFID reader emerges from it beyond this, the tester extends 14 cm backwards beyond the platform. The size of the platform itself (for tester version with stand): 500 x 490 x 20 mm.
Tilt angle of the stand: 5 degrees

sizes

Mounting plate sizes

Tartólemez méretekTartólemez méretek

REQUEST FOR QUOTATION

ORDER CODE TYPE
AIJGO-34 Personal grounding tester with dual foot electrode, RFID reader, operating software, mounting plate and universal holder {AIJGO-34}
AIJGO-STA Stand for AIJGO tester {AIJGO-STA }
AIJGO-34/MSW PieManager management software for AIJGO-34 and AIJGO-34/COMP (if buying more than one tester, it is sufficient and recommended to buy one if they will be networked together) {AIJGO-34/MSW
}

 Aijgo logo

Partners